Single Sheet Tester SST 500X500

Model: SST500X500
Gopal Make single sheet tester SST is developed as per IEC 60404-3. It can be attached as probe to Ep400 Main unit and user can test specific total loss, specific apparent power, and r.m.s value of magnetic field strength as well as B/h curve in the software.
 
The single sheet tester with the test specimen represent as unloaded transformer, the total loss of which is measured by use of flux coil and two U Shape precision low loss yoke. User friendly mechanism is provided to up lift the upper yoke and to easily insert the sample in between two yoke and winding. Testing is done automatically by using Ep400 main unit & software
 
SST Probe is optional and not included in Ep400 main unit. 

Gopal electronics has built the reputation in the market on basis of transparency in sales, easy and quick service, and economic and reliable rates for repairing.

  • The whole product range of Gopal Electronics comes with standard 1 years warranty.
  • For easy warranty claim, you can register your product on our website in case you purchased from any dealer.
  • You can send your instrument by courier to our works for any repairing or calibration service. Your instrument will be inspected first and you will be notified for its repairing or calibration charges. The repairing or calibration process only carried out after the confirmation of customer.
  • The time for repairing or calibration is maximum 72 hours after confirmation.

Calibration Service

We provide traceable calibration service according to ISO17025 for all of our product range. Every year you require to calibrate your instruments for your factory certification like ISO, and to maintain quality of your routine test and quality inspection.

 

Calibration by Gopal Electronics:

  • Our reference meters and standards are traceable to NABL accredited lab or our internal standards as per ISO17025.
  • We have calibration facility of 0.007% (66 ppm) energy accuracy for sinusoidal waveforms.
  • As manufacturer we can change the calibration data and compensate the deviation in each measuring point. So you can get actual calibration not only the comparison report.
  • General Service, replacement of minor parts can be carried out at the same time.
  • Free firmware upgrade is also the part of our calibration process.
  • We can provide calibration at special points which are not included in usual calibration on customer request, like calibration at 0.1 power factor.
  • It takes maximum two to four days to complete the calibration process after the receipt of your instrument at our factory.
Cancel
  • complies to Internation standard IEC 60404-3
  • Robust mechanical structure
  • Low Loss Yoke
  • Special Mechanisum for uplift upper yoke to insert sample
  • Specific total loss of yoke less than 1watt/kg @1.5T 50Hz
  • Sample size 500mm x 500mm
  • The Single sheet tester is applicable to test specimens obtained from magnetic sheets and strip of any quality. The magnetic characteristics are determined form a sinusoidal induced voltage, for specific peak value of magnetic polarization and for a specific frequency 
  • User can measure specific total loss, specific apparant power, and R.M.S value magnetic field strength as well as B/H curve in software
  • Minimum Sample size = 500mm x 300mm
  • Standard Sample size = 500mm x 600mm
  • Weight of Probe Unit = Approx. 300kg 
  • Specimen weight    = Maximum 1kg
  • Dimension     = Approx L62 x W42 x H65 inch
  • complies to IEC 60404-3
  • Specific total loss of yoke less than 1watt/kg @1.5T 50Hz
  • Sample size 500mm x 500mm
  • Main SST Probe 500x500
  • SST connection wire set